Item Title: Automatic External Defibrillator Philips Heart Start Frx
Item Description: Defibrillator contents :
It should Include defibrillator, battery, carry case, smart pads II (1 pre connected set, 1 spare set), setup and maintenance guides, owner-s manual, Quick reference guide, data sticker
Waveform:
It should be Biphasic truncated exponential: wave form parameters are automatically adjusted as a function of patient defibrillation impedance
Therapy:
Adult defibrillation:
Nominal peak current 32 A (150 J nominal into a 50 ohm load).
Pediatric defibrillation with optional Infant/child key installed:
Nominal peak 19 A (50 J nominal ohm load)
Protocol:
Device should follow preconfigured settings; defibrillation and CPR protocol should be customized using heart start event review pro software.
Sealing:
Should have water jet-proof IPx5 per IEC60529
Dust-protected IPx5 per IEC60529
Altitude:
0 to 4, 572 m (0 to 15000ft)
Aircraft:
Should meet RTCA/DO-160 Section 21 (Category M - Radiated Emissions) and Section 20 (Category M - Conducted Immunity, and Category D - Radiated Immunity)
Crush:
500 kg (1100 lb)
Drop:
Should Withstand 1.22 m (4 ft) drop on any edge, corner or face of the device onto masonry surface
EMI (radiated/ immunity):
Should meet CISPR 11 Group 1 Class B and IEC 61000-4-3
Item Quantity: 1
Unit of Measure: NUMBERS
Delivery Period (In number of days): 15