Expired Tender Ghana ICB TOT Ref. 91135745

Supply of Medical Equipment Tender in Ghana - 91135745

Tender notice from Ghana, published on 26 Oct 2023.

Executive Summary

The DEUTSCHE GESELLSCHAFT FÜR INTERNATIONALE ZUSAMMENARBEIT has issued a Tender notice for the procurement of a Supply of Medical Equipment in the Ghana. This Tender notice was published on 26 Oct 2023 and is scheduled to close on 31 Oct 2023, with an estimated Tender value of Refer Document. Interested bidders can access detailed Tender information, eligibility criteria, and complete bidding documents by referencing TOT Ref No. 91135745, while the tender notice number is 12.1003.8-439.01 and Register on the platform.

Key facts

Summary
Supply of Medical Equipment
TOT Ref. No.
91135745
Document Ref. No.
12.1003.8-439.01
Country
Ghana
Notice type
Tender
Competition
ICB
Publication date
26 Oct 2023
Bid deadline
31 Oct 2023
EMD
Refer Document
Tender value
Refer Document
Purchaser ownership
Public
Financier
Deutsche Gesellschaft für Internationale Zusammenarbeit (GIZ) GmbH

Purchaser details

Purchaser contact is for registered users

Login to see details

Tender details

Tenders are invited for Supply of Medical Equipment
CENTRIFUGE

1. This is a Laboratory centrifuges to be used in chemistry,

biology, and biochemistry for isolating and separating solids

from liquids in a suspension.

2. Brushless induction motor centrifuge for a Hospital

Laboratory

3. Will be used in the laboratory for separating components of

body fluids of different densities.

4. Speed (RPM) should not be greater than ±5% of the

displayed speed.

5. The centrifuge should have a minimum of 12 place rotor for

2ml to 10ml tubes

6. Must be supplied with the test tube adapter where necessary

7. The centrifuge should have a programmable digital timer.

It is preferred that the timer be adjustable in one-minute

increments.

8. The timer should have a timing range of at least one hour.

9. The centrifuge will have an alert indicator for imbalance,

current, and lid open.

10. The centrifuge should have an alert indicator when cycles

completely stop.

11. The centrifuge should not operate with lid open. It is

preferred that the centrifuge have a lid interlock.

12. The proposed centrifuge should have the following

parameters:

a. Maximum speed (minimum 8000 ±10% RPM)

b. Rotational centrifugal force (RCF) (minimum 2086g)

c. Type of rotor head (swinging bucket)

d. Braking type should be electric

e. Display type digital

f. RPM or rotor calibration method (e.g., tachometer)

13. It should be supplied with instruction for use and technical

manuals in English

Standard: CE and ISO certified

Please Indicate Brand Name and Model Number

1PCS

Continued on next page...

2

9. It should have 4 temperature presets

10. The temperature stability accuracy of the water bath should

be ±0.1°C

11. The temperature uniformity of the water bath should

be±0.2°C

12. It should have dimensions of: (L x W x H) 9.7 x 14 x 9.1 in.

(246 x 355 x 232 mm)

13. It should fulfill the electrical requirements of 120/230V

50/60 Hz

14. It should satisfy an average amperage of 2.2-2.7 / 1.0-1.3

(120V/230V)

Please Indicate Brand Name and Model Number

1pcs

3

LABORATORY MICROSCOPE

1. The light microscope should consist of the lens system,

body, stage and stage controls and Illumination system.

2. The observation tube should be Siedentopf with ergonomic

inclination binocular tubes with adjustable inclination of

between 5-35°

3. The power of the eyepieces should be 10x

4. The eyepiece should be achromatic widefield type with high

eyepoint

5. The interpupillary distance adjustment range should

be approximately 50 to 75mm to ensure comfortable

observation condition for every user

6. The binocular observation tube should be equipped with a

dioptre adjustment

7. The revolving nosepiece should be tilted inward and must

be configured for 5 positions

8. The objectives should be achromatic with universal infinity

corrected optics. Other features such as universal plan

fluorite, plan achromatic, plan apochromatic, universal plan

achromatic

9. The minimum magnification range of the objective lens

should be from 2x to 100x including oil immersion

10. Contrast methods should include, phase, brightfield,

darkfield, epifluorescence, polarization

11. The Illumination system should consist of a light source

provided by an LED light source or alternatively with 6V

30W halogen bulb with variable intensity and provided with

a separate on/off switch. The lifespan of the mandatory

bulb(s) together should last for not less than 10, 000hrs

12. The light source should have an emergency backup system

in event of bulb failure.

13. The total magnification should range from 20x to at least

1000x.

14. The Stage should be mechanical, graduated and ceramic

coated with a coaxial tension adjustment with stage

dimension approximately 188(W) x 134(D)mm moving in

both X 76mm and Y 50mm direction.

15. The scale graduation shall be in white on a black background

with 1µm increments.

16. The stage shall include coaxial X-Y controls, and motion

should be roller-bearing.

17. The supply of the microscopes should include the following

18. Durable dust cover, set of spare bulbs, bottle of at least 25

ml immersion oil, a roll of lens tissue paper, 100 ml of lens

cleaning solution, one anti-static cleaning brush, 2 sets of

fuses if required

19. Standards: CE certified

Please Indicate Brand Name and Model Number

2PCS

2

WATER BATH

1. The chamber material should be Stainless Steel

2. The water bath should be a general purpose with a capacity

of 5L

3. The machine must be CE and EU certified

4. It must have a temperature range of ambient temperature

to 100°C

5. The display should be a Monochrome LCD

6. The water bath should have a heating capacity 300W

7. It should include clear Polycarbonate Gable Cover and

Rubber Duck

8. The chassis of the water bath should be made of Epoxy

Powder-coated Steel

ELECTROPHORESIS MACHINE

1. This is an electrophoresis equipment for separating

molecules (such as proteins, DNA, or RNA fragments)

according to size and electrical charge by applying an

electric current to them. Each kind of molecule travels

through the medium at a different rate, depending on its

electrical charge and molecular size.

2. This should be a complete system for rapid electrophoresis

of proteins & nucleic acids for the diagnosis of Sickle Cell

Disease and other haemoglobinopathies

3. The unit should be horizontal electrophoresis apparatus

made from thick injection-moulded UV transparent chamber

base and cover

4. It should be durable and leak-proof

5. The lid should include a silicon rubber gasket and must fit

securely in place but should be easy to remove

6. Should have adjustable levelling feet, recessed electrical

wires, and a safety stop to prevent the gel from running

when the cover is not securely fitted to the unit.

7. Should be able to run two slab gels 10 x 8 cm approximately

simultaneously

8. Complete gel casting system for casting multiple gels

9. Each cast tray with embossed gel tray, ruler should have

rubber end caps and multiple slots for easy and versatile

comb placement.

10. Easy-to-read digital display.

11. Safety features for overload, sudden load change, short

circuit protection etc. and personal and environmental

protection.

12. Automatic recovery after power failure.

13. Should be supplied with all the required accessories and

spare parts.

14. Should operate satisfactorily on power input of 200-260VAC,

50Hz and fitted with British Standard plug with cable length

of 3m.

15. Should be equipped with resettable overcurrent breaker for

protection

Should be FDA and/or CE approved product

Please Indicate Brand Name and Model Number

1PCS

5

ADULT WEIGHING SCALE

1. This is the dial type of scale for weighing patients and

determines their height to assess their nutritional status.

2. The weighing scale should include Height Measure and

Body-Mass Index (BMI) scale

3. The weighing scale shall have a weighing range of 0.0kg to

200kg.

4. The weighing scale shall have a readability of 0.5kg.

5. The weighing scale shall offer metric calibrations.

6. Should have a waist high, large, and easy to read dial (“Big

Ben clock”) type of analogue display

7. The low-level platform should have a slip resistant cover

8. The integrated height measure should be telescopic with

eye level readout

9. The height measure should have maximum height of

approximately 230cm with 1mm graduation

10. The BMI scale should be colored coded

11. User manual in English (should be included in the offer)

12. Standards: CE certified

Please Indicate Brand Name and Model Number

1PCS

6

PHARMACY FRIDGE

1. Refrigerator specially designed to store pharmaceuticals,

medicine, or live vaccines.

2. Approximate capacity of at least 20 Cu. Ft. or 500 litres

3. With a single, lockable, stainless-steel door with glass window which can be hermetically sealed

4. Door with self-closing system and locking without mouldings or inside shelves

5. Thermal insulation of CFC-free polyurethane foam

6. CFC-free refrigerant

7. Operating temperature: +2 to +8 ° C

8. Temperature uniformity of +/- 1°C

9. Forced air recirculation

10. Manual or automatic defrosting system

11. Automatic condensate evaporation

12. Microprocessor controlled temperature with digital display

indicator

13. Visual (and audible) alarm system to indicate failure in

the power supply, temperatures above or below the programmed and sensor failure

14. Interior with adjustable drawers and racks made of stainless steel

15. With interior lighting for door opening

16. Powder-coated exterior and cabinet

17. Energy efficiency label is at least the second highest of the

current guidelines

Dimensions are minimum. (WxDxH): 700 x 900 x 2000 mm

Please Indicate Brand Name and Model Number

1PCS

7

MICROSCOPE SLIDES

1. Slide size: 1” x 3” (25.4x76.2mm)

2. Material: optical glass

3. Double Frosted Ends

4. Ground Edges

5. Pre-cleaned 50pc microscope blank slides in a box

5PCS

8

MICRO PIPETTE

1. This is a single channel and handheld adjustable volume

pipettes, for dispensing fluid in the clinical laboratory

2. The pipette shall have a distinctive handle with ejector

pushbutton located on the top of the finger rest with the

volume counter facing backwards to offer users comfort and

continuous control of the pipette volume

3. The unit shall be supplied with pipettes tips which shall be

easy to remove

4. The tips shall be separate from the pipette mechanism to

prevent cross-contamination.

5. The pipette shall safely dispense a wide range of volumes

ranging from at least 1.0-10 µl; 10-100 µl; 100-1000 µl;

all with accuracy of approximately ±0.5% and precision of

approximately ±0.2%.

6. Must be supplied with corresponding pipette tips which

shall be either colour coded by size or clearly identified by

volume sizes to make identification easy.

7. The pipettes should be made of corrosion proof material

and sho

Documents

Tender Notice
Login to view

Ghana Public Procurement Intelligence

Frequently asked questions

The bid deadline for Supply of Medical Equipment was 31 Oct 2023. Bidding is closed.

This tender has been issued by DEUTSCHE GESELLSCHAFT FÜR INTERNATIONALE ZUSAMMENARBEIT in Ghana.

The tender value is not disclosed; please refer to the tender document. EMD / bid security details are provided in the tender document.

This tender is financed by Deutsche Gesellschaft für Internationale Zusammenarbeit (GIZ) GmbH.

The tender notice and related documents are listed in the Documents section above. Select a document to view or download it. Please log in or register to access the documents. If you cannot find additional documents in the Documents section, please log in to your account and Send Document Request from your account section.

The TendersOnTime reference is 91135745. The purchaser document reference is 12.1003.8-439.01.