Expired Tender Hong Kong ICB TOT Ref. 62988887

Supply and Installation of Radiographic Tender - 62988887

Tender notice from Hong Kong, published on 09 Feb 2022.

Executive Summary

The HOSPITAL AUTHORITY has issued a Tender notice for the procurement of a Supply and Installation of Radiographic / Fluoroscopic Systems. Cardiovascular for Division of Cardiology, Department of Medicine of New Block Redevelopment Project. Queen Mary Hospital in the Hong Kong. This Tender notice was published on 09 Feb 2022 and is scheduled to close on 09 Mar 2022, with an estimated Tender value of Refer Document. Interested bidders can access detailed Tender information, eligibility criteria, and complete bidding documents by referencing TOT Ref No. 62988887, while the tender notice number is HKWC-T026/20(P-QMH) and Register on the platform.

Key facts

Summary
Supply and Installation of Radiographic / Fluoroscopic Systems. Cardiovascular for Division of Cardiology, Department of Medicine of New Block Redevelopment Project. Queen Mary Hospital
TOT Ref. No.
62988887
Document Ref. No.
HKWC-T026/20(P-QMH)
Country
Hong Kong
Notice type
Tender
Competition
ICB
Publication date
09 Feb 2022
Bid deadline
09 Mar 2022
EMD
Refer Document
Tender value
Refer Document
Purchaser ownership
-
Financier
Self Financed

Purchaser details

Purchaser contact is for registered users

Login to see details

Tender details

Expression of Interest are invited for Supply and Installation of Radiographic / Fluoroscopic Systems. Cardiovascular for Division of Cardiology, Department of Medicine of New Block Redevelopment Project. Queen Mary Hospital. Hong Kong West Cluster. Hospital Authority
1.1 This exercise calls for the supply and installation of Two

sets of X-ray generating system shall be fully equipped for

Monoplane and Bi-plane cine fluorography and digital

imaging Operations for New Block Redevelopment Project

of Queen Mary Hospital, Hong Kong West Cluster,

Hospital Authority. The tentative completion of

construction is by 3Q 2024.

(M)

1.2 There are two items of product required in the EOI

exercise (refer to Appendix I for details):

Item 1: X-ray generating system with Anesthesia Pendant

Item 2: X-ray generating system with Cardiologist Pendant

(M)

1.3 The offer will be considered and accepted on an overall

basis.

(M)

Part A Main System for item 1 and item 2 (Clause 2 - 12)

2 General

2.1 High frequency or constant potential DC output. (M)

2.2 Microprocessor controlled. (M)

2.3 Solid-state rectification. (M)

2.4 Maximum power output not less than 100kW. (M)

Page 11 of 171

EOI Ref.:HKWC-T026/20(P-QMH)

Clause Specification M/D/O

Yes No

(Please tick and provide

details as appropriate)

2.5 Maximum continuous power output: 2.5kW for 15 minutes

or better.

(M)

2.6 Maximum current shall not be less than 1000mA at 80kV (M)

2.7 Radiography kV range shall be 50kV or lower up to

125kV or higher.

(M)

2.8 Fluoroscopy kV range shall be 60kV or lower up to 110kV

or higher.

(M)

2.9 Suitable for monoplane and bi-plane pulsed fluoroscopy

connection to not less than 2 dual-focus/tri-focus X-ray

tubes.

(M)

2.10 Automatic line voltage compensation. (M)

2.11 Automatic exposure control. (M)

2.12 Support pulsed X-ray of up to at least 25 frames/s in single

plane

and biplane mode for digital dynamic acquisition at

1024x1024 or better digital matrix.

(M)

2.13 Tube overloads indication and/or protection. (M)

2.14 Include not less than 50 selectable programs for LV

angiography, coronary angiography, pediatrics

angiography etc. to avoid waiting time during the

procedure.

(M)

2.15 Preference will be given to highest selectable programs.

Please specify your offer:

(D)

3

X-ray Tubes

3.1 Two X-ray tubes, at least dual focus spot mounted to the

imaging Equipment Support Systems Frontal Arm and

Lateral Arm with the following features:

(M)

Page 12 of 171

EOI Ref.:HKWC-T026/20(P-QMH)

Clause Specification M/D/O

Yes No

(Please tick and provide

details as appropriate)

3.2 The x-ray tubes shall operate in continuous high-speed

rotation of the anode with special lubricant bearing

mechanism.

(M)

3.3 Focal spot sizes (mm) in the range of not more than 0.5 for

small focus and not more than 1.2 for large focus spot.

(M)

3.4 Grid Switching at pulsed fluoroscopy or equivalent.

Tenderer shall provide details: _____________________

(M)

3.5 Maximum anode heat dissipation in continuous mode shall

be at least 3.5kW or better.

(M)

3.6 Maximum tube voltage in kV shall not be less than 125kV. (M)

3.7 Maximum tube loading in kW shall not be less than 80

kW.

(M)

3.8 Maximum heat storage capacity in X-ray tube housing unit

shall not be less than 5.0 MHU.

(M)

3.8.1 Preference will be given to the offer with highest heat

storage capacity in X-ray tube housing unit. Tenderers

shall quote the value for reference.

(D)

3.9 The maximum cooling rate of the anode shall not be less

than 900 kHU/min.

(M)

3.9.1 Preference will be given to the offer with highest

Maximum continuous cooling capacity of the anode.

Tenderers shall quote the value for reference.

(D)

3.9.2 Maximum assembly continuous heat dissipation 4000W or

better

(D)

3.10 Provision of a dose reduction filter (operator selectable or

automatic) for filtering of low energy radiation to optimize

image quality and dose efficient.

(M)

3.11 Direct anode oil cooling system, or equivalent. (M)

3.12 Radiation leakage-proof standard of the tube housing shall

conform to the current ICRP requirements and the

Radiation Ordinance of Hong Kong.

(M)

3.13 Cumulative patient dose measurement via dose area

product measurement.

(M)

3.14 In addition to the above-specified items, Tenderer shall

provide the price and terms of guarantee for the X-ray

tube.

(M)

3.15 Maximum anode heat storage capacity shall not be less

than 6.0 MHU.

(D)

3.16 Maximum current in acquisition shall not be less than

1000mA

(M)

4

Beam-limiting device

4.1 Vertical and horizontal shutters. (M)

4.2 Anatomical filters for application of digital subtractions. (M)

4.3 Independent rotation of the anatomical filters. (M)

4.4 Copper filters shall be use for dose reduction. (M)

4.5 The thickest copper equivalent filter shall be 0.9mm or

thicker.

(M)

4.5.1 Preference will be given to the offer with 1mm or thicker

copper equivalent filter. Please specify your offer:

(D)

4.6 Automatic limitation of the X-ray beam to the size of the

selected input field of the flat panel detector in both

fluoroscopy and digital acquisitions. The selected input

field size of the exposure area shall be user adjustable.

(M)

4.7 Allows virtual collimation and adjustment of anatomical

filters with reference to Last-Image-Hold (LIH) without

the need of further fluoroscopy.

(M)

8

Image Displays System

8.1 Two sets of medical-grade large ceiling suspended high

definition, DICOM compatible color LCD/LED display in

the Examination Room with the following features:

(M)

8.2 Display with size no less than diagonal 58” (M)

8.3 Native resolution: 3840x2160 pixels or better (M)

Page 19 of 171

EOI Ref.:HKWC-T026/20(P-QMH)

Clause Specification M/D/O

Yes No

(Please tick and provide

details as appropriate)

8.4 Brightness: 450 cd/m2 or better (M)

8.5 Contrast ratio: 4000:1 typical, or better. (D)

8.6 Capable of high contrast display for live, reference for both

imaging planes and roadmap images

(M)

8.7 Capable to offer viewing of total not less than 8 sources

simultaneously including HD grayscale medical images,

prior image comparison, real time ultrasound images,

reconstructed or multi-modality color images and patient

monitoring

(M)

8.8 Resize and/or enlarge information at any stage during the

case

(M)

8.8.1 The size of each windows on the screen is freely adjustable (M)

8.9 Module for screen layout control at tableside (M)

8.10 Screen layout module allows the user direct and switches

the video output of all connected medical equipment to

specific sub windows of the big monitor

(M)

8.11 One set of monitor shall mount on a height-adjustable

ceiling suspended display carriage inside the Examination

room. The other set shall be mounted on wall.

(M)

8.12 Two medical grade monitors shall be provided and mount

on the ceiling suspended display carriage for displaying

live image (lateral and frontal) for contingency

(M)

8.13 The medical grade monitors shall be of at least 24"

(diagonal)

(M)

8.13.1 One (1) set of display carriage (M)

8.13.2 Ceiling-suspended carriage for the video images displays (M)

8.13.3 Longitudinally movement shall not less than 300cm (M)

8.13.4 Rotation of the ceiling-suspended support axis shall not be

less than 300˚

(M)

Page 20 of 171

EOI Ref.:HKWC-T026/20(P-QMH)

Clause Specification M/D/O

Yes No

(Please tick and provide

details as appropriate)

8.13.5 It shall allow height adjustment. (M)

8.13.5.1 The control of the height adjustment shall be motorized. (M)

8.14 Image Displays in Control Room (LCD/LED monitors

shall be provided)

(M)

8.14.1 Two (2) medical-grade of at least 24 inch LCD/LED

display monitors for the live & reference images (first

monitor showing frontal live and lateral live image, second

monitor showing frontal reference and lateral reference

image)

(M)

8.14.2 One (1) additional at least 24 inch medical-grade flat-panel

display shall also serve as a display for the Control

Console.

(M)

8.15 Provision of an integrated video switch/connection box for

high quality video sources.

(M)

9

Radiographic Table

9.1 One floor-mounted angiographic table with the following

features:

(M)

9.2 Tabletop shall be made of material of low X-ray absorption

material, properly shaped to the human body and without

shadow-producing and construction parts from head to

foot.

(M)

9.3 4-way floating tabletop non-transparent with electromagnetic brakes for tabletop movements.

(M)

9.4 Hand switch to be provided for releasing brakes of tabletop

movements.

(M)

9.5 Total longitudinal tabletop adjustment to be not less than

1200mm.

(M)

9.6 Total transverse tabletop adjustment to be not less than

350mm.

(M)

Page 21 of 171

EOI Ref.:HKWC-T026/20(P-QMH)

Clause Specification M/D/O

Yes No

(Please tick and provide

details as appropriate)

9.7 Tabletop shall have a motorized height adjustment in a

range of 780 mm - 1000 mm or wider above floor level.

(M)

9.7.1 Table positions can be stored and recalled from touch

screen module, and this can be done independent without

affecting lateral and frontal arm position.

(M)

9.8 Table titling range shall be -15 degrees to +15 degrees or

wider.

(M)

9.9 Tabletop shall have a pair of side rails for the attachment

of accessories.

(M)

9.10 Table base or mounting column to incorporate connection

panel for footswitch, biplanes control, and power line

socket.

(M)

9.11 Maximum patient weight shall not be less than 200kg. (M)

9.12 Native table-top length: at least 260cm. (M)

9.13 To be provided with the following accessories:

9.13.1 Two handgrips (M)

9.13.2 Pan Handle for floating movements of the table top (M)

9.13.2.1 Can operate tabletop brake release (M)

9.13.3 One set of latex free mattress with 7 cm thickness and it

should adapt to the shape of the patient-s body. (M)

9.13.4 Two sets of arm support boards. (M)

9.13.5 Two sets of elbow

Documents

Tender Notice
Login to view

Hong Kong Public Procurement Intelligence

Frequently asked questions

The bid deadline for Supply and Installation of Radiographic / Fluoroscopic Systems. Cardiovascular for Division of Cardiology, Department of Medicine of New Block Redevelopment Project. Queen Mary Hospital was 09 Mar 2022. Bidding is closed.

This tender has been issued by HOSPITAL AUTHORITY in Hong Kong.

The tender value is not disclosed; please refer to the tender document. EMD / bid security details are provided in the tender document.

This tender is Self Financed

The tender notice and related documents are listed in the Documents section above. Select a document to view or download it. Please log in or register to access the documents. If you cannot find additional documents in the Documents section, please log in to your account and Send Document Request from your account section.

The TendersOnTime reference is 62988887. The purchaser document reference is HKWC-T026/20(P-QMH).

Browse other tenders from Hong Kong or search by relevant keywords and CPV classification.