Supply and Installation of Analyzers, Laboratory Tender - 68429354
Tender notice from Hong Kong, published on 01 Jul 2022.
Executive Summary
The HOSPITAL AUTHORITY has issued a Tender notice for the procurement of a Supply and Installation of Analyzers, Laboratory, Hematology, Cell Counting in the Hong Kong. This Tender notice was published on 01 Jul 2022 and is scheduled to close on 08 Aug 2022, with an estimated Tender value of Refer Document.
Interested bidders can access detailed Tender information, eligibility criteria, and complete bidding documents by referencing TOT Ref No. 68429354, while the tender notice number is KEC/ADM/CONF/PFPE/13/005/2223 and Register on the platform.
Key facts
Summary
Supply and Installation of Analyzers, Laboratory, Hematology, Cell Counting
Tenders are invited for Supply and Installation of Analyzers, Laboratory, Hematology, Cell Counting, Automated for Pathology Department for Expansion of United Christian Hospital, Kowloon East Cluster, Hospital Authority 1.1 This tender calls for the supply, delivery, installation and at least 12-month warranty of one set of Haematology
Laboratory Cell Counting Automation System at the Haematology Laboratory at United Christian Hospital
(UCH).
(M)
1.2 Tenderer shall guarantee that all analytical components, consumables, and reagents are manufactured in
accordance with approved international standards, and those materials supplied shall be United States- Food and
Drug Administration (FDA) approved and CE marked. Only CE mark is not acceptable.
(M)
1.3 The Haematology Laboratory Cell Counting Automation System for UCH shall have:
1.3.1 The system shall have pre- and post- analytic tube management functions. (M)
1.3.2 No more than six units of identical haematology blood cell analyzers with total throughput of not less than 600
samples per hour for Complete Blood Count (CBC) with White Blood Cell (WBC) Differential.
(M)
1.3.3 Not more than 3 units of automatic slide maker/stainers shall have a total throughput of not less than 150 slides
per hour.
(M)
1.3.4 One digital cell morphology analyzer with total throughput of not less than 30 samples per hour to locate and to
display images of blood cells on peripheral blood smears.
(M)
1.3.5 A diluent preparation unit that can prepare diluent from concentrated diluent with laboratory grade water for
performing CBC shall be provided if required. It shall include storage of the prepared diluent and shall have an
alert system to monitor the quality of prepared diluent.
(M)
1.3.6 Blood cell analyzers, slide maker/stainers ; and reagent reconstitution systems if applicable, connected in one
complete sample transportation system functionable at the time of delivery and installation.
(M)
1.3.7 Tube sorter, blood cell analyzers, slide maker/stainers, digital cell morphology analyzer, and reagent
reconstitution systems connected in one complete sample transportation system functionable at the time of
delivery and installation.
(D)
1.3.8 One data management middleware, which shall consolidate and manage the results from different blood cell
analyzers, slide maker/stainers, and digital cell morphology analyzer.
(M)
1.3.9 The system shall be of floor stand type. Tenderer shall also provide sufficient number of working benches or
desks to accommodate the controllers and their associated workstations and printers. The benches or desks shall
be aesthetically and ergonomically designed.
(M)
1.3.10 Dimension of the whole configuration shall fit with the composite drawing as Appendix A: (M)
1.3.11 Dimension of the whole configuration, which shall fulfill clause A1.3.9, and shall confine to the space available
with dimension of 8, 000mm (W) x 2, 000mm (D) x 1, 500mm (H).
(M)
B Sample Transportation System
1 The configuration of the sample transportation system shall be flexible, linear design, to fit the floor plan at
UCH.
(M)
1.1 Tenderer shall quote for an optional item which enable the configuration of the sample transportation system to
turn at 90° if available. Details shall be provided in the Tender return for consideration. Tenderer shall quote
the unit price in Schedule A (Price) - Section E.
(O)
2 The track shall be expandable and can be connected to not less than 9 analyzers, for future upgrade or to serve
for workload increase purpose. The track shall have the capability to connect to Hemoglobin A1c (HbA1c) and
Erythrocyte Sedimentation Rate (ESR) system if necessary.
(D)
Page 33 of 167
Tender Ref.: KEC/ADM/CONF/PFPE/13/005/2223 (To be included in the Technical Proposal)
Company Chop: ______________________
Clause Requirement M/D/O Yes No
(Provide Details)
3 It shall allow the sample to perform re-run, repeat and reflex within the same haematology analyzer or between
haematology analyzers according to user pre-defined rules.
(M)
3.1 The reflex function for platelet should be able to trigger the analyzer to switch to a more advanced methodology
automatically; when impedance method is affected by interference, fluorescent measurement of Platelet Count
(PLT) could be employed.
(D)
4 It shall comprise of additional lane that allows the sample rack to route onto the by-pass lane while other lanes
are occupied.
(M)
5 The sample loading and collecting yards shall be placed together on the same side. (M)
6 The sample loading and collecting yards can be placed separately on the opposite side of the track to allow a
flexible workflow design.
(M)
7 The capacity of the sample loading yard and collecting yard shall each accommodate not less than 250 samples
to allow sample continuous loading.
(M)
8 The cell counting automation system shall be able to connect with existing cell counting system (Brand &
Model: Sysmex XN9100) in one integrated system.
(D)
9 The track system shall have the following functions:
9.1 The pre-analytic system shall be capable of:
(i) Continuous loading of sample tube
(ii) Identifying sample tubes by barcode labels
(M)
9.2 The pre-analytic system is desirable to be capable of:
(i) Sorting sample tube out for offline analysis
(ii) Sorting error sample (e.g. unidentified barcode label, invalid test order) tube out at a designated area
(iii) Prioritizing urgent sample by rack type
It is desirable to be able to sort post-analytical tubes for sample management base on user-s preference, and store
or archive no less than 1000 tubes.
(D)
9.3 It can be placed at the sample entry point, between the haematology analyzers and the slide-maker stainers, or
at the end of the track system to tailor-fit the lab order profiles and sample workflow.
(M)
10 The system shall use the waste drainage point indicated in Appendix A. The rear panel shall face the drainage
system to ensure heat dissipation does not affect the work benches. The successful tenderer shall provide the
drain pipe for the waste water and discharge to the waste drainage point. The successful tenderer shall provide
an exhaust air system to cater the heat dissipation, if necessary.
(M)
11 The successful tenderer shall notify the Main Contractor for the required size of the waste drainage point for the
equipment.
(M)
12 The successful tenderer shall notify the Main Contractor for the location and size of the external wall/ slab
openings and the weatherproof louvre size for the exhaust air outlet if the exhaust air system to be provided by
themselves.
(M)
13 The sample entry point of the system shall be placed closest to the laboratory entrance in Appendix A to reduce
samples transportation time. The sample collection point shall be placed on the opposite side.
(M)
14 Sufficient storage racks for pre- and post-analytical unit shall be available to accommodate no less than 10, 000
CBC samples.
(M)
C Haematology Blood cell Analyzer Unit
1 The analyzers shall employ the following measuring technologies for CBC, WBC differential (DIFF) and
reticulocyte (RET) analysis:
Page 34 of 167
Tender Ref.: KEC/ADM/CONF/PFPE/13/005/2223 (To be included in the Technical Proposal)
Company Chop: ______________________
Clause Requirement M/D/O Yes No
(Provide Details)
1.1 Fluorescence flow cytometry method using laser diode for WBC, WBC differential and reticulocyte
measurement to allow maximum efficiency with not less than 70 samples per hour for CBC+DIFF+RET per unit
of analyzer.
(D)
1.2 Impedance detection method for RBC measurement. (M)
1.3 Direct impedance measurement method for RBC Mean Corpuscular Volume (MCV). (D)
1.4 Cyanide-free colorimetric method for hemoglobin measurement (M)
2 The analyzers shall employ the following analysis mode for different samples analysis:
2.1 Whole blood mode (M)
2.2 Pre-Dilution mode (M)
2.3 Low WBC mode (M)
2.4 Body fluid mode (M)
3 The analyzers shall employ the following discrete mode for measurement:
3.1 CBC only (M)
3.2 CBC/ DIFF (M)
3.3 CBC / DIFF / RET (M)
4 The analyzers shall be able to provide CBC, 5-part WBC Differential, Reticulocyte and nucleated red blood cell
count on whole blood sample. Reportable parameters shall include:
(M)
4.1 White Blood Cell Count (WBC) (M)
4.2 Red Blood Cell Count (RBC) (M)
4.3 Neutrophil Percentage (NEUT%) (M)
4.4 Neutrophil Count (NEUT#) (M)
4.5 Lymphocyte Percentage (LYMPH%) (M)
4.6 Lymphocyte Count (LYMPH#) (M)
4.7 Monocyte Percentage (MONO%) (M)
4.8 Monocyte Count (MONO#) (M)
4.9 Eosinophil Percentage (EO%) (M)
4.10 Eosinophil Count (EO#) (M)
4.11 Basophil Percentage (BASO%) (M)
4.12 Basophil Count (BASO#) (M)
4.13 Nucleated RBC Percentage (NRBC%) (M)
4.14 Nucleated RBC Count (NRBC#) (M)
4.15 Hemoglobin (HGB) (M)
4.16 Hematocrit (HCT) (M)
4.17 Mean Corpuscular Volume (MCV) (M)
4.18 Mean Corpuscular Hemoglobin (MCH) (M)
4.19 Mean Corpuscular Hemoglobin Concentration (MCHC) (M)
4.20 RBC Distribution Width-CV (RDW-CV) (M)
4.21 RBC Distribution Width-SD (RDW-SD) (M)
4.22 Platelet Count (PLT) by 3 different methods:
4.22.1 By Impedance method (M)
4.22.2 By Optical method and by fluorescence method with platelet-specific fluorescent stain (D)
The bid deadline for Supply and Installation of Analyzers, Laboratory, Hematology, Cell Counting was 08 Aug 2022. Bidding is closed.
This tender has been issued by HOSPITAL AUTHORITY in Hong Kong.
The tender value is not disclosed; please refer to the tender document. EMD / bid security details are provided in the tender document.
This tender is Self Financed
The tender notice and related documents are listed in the Documents section above. Select a document to view or download it. Please log in or register to access the documents. If you cannot find additional documents in the Documents section, please log in to your account and Send Document Request from your account section.
The TendersOnTime reference is 68429354. The purchaser document reference is KEC/ADM/CONF/PFPE/13/005/2223.
Browse other tenders from Hong Kong or search by relevant keywords and CPV classification.