Procurement Summary
Country: Philippines
Summary: SUPPLY AND DELIVERY OF OPTHALMOLOGY EQUIPMENT FOR USE OF GEN. EMILIO AGUINALDO MEMORIAL HOSPITAL.
Deadline: 01 Dec 2015
Posting Date: 18 Nov 2015
Other Information
Notice Type: Tender
TOT Ref.No.: 4364956
Document Ref. No.: gleyva-G255-2015
Competition: ICB
Financier: Self Financed
Purchaser Ownership: -
Tender Value: Refer Document
CPV Classification
33122000 - Ophthalmology equipment
Purchaser's Detail
Name: Login to see tender_details
Address: Login to see tender_details
Email: Login to see tender_details
Login to see detailsTender Details
1 unit Vision Enhancement System
1 unit Auto Kerato Refractometer Interlocal Health Zone
PHACOEMULSIFICATION MACHINE
Specifications:
Dimension (L x W x H)
457 x 457 x 1220mm
Weight max 144kg.
Power Supply
Voltage 100-240V
Frequency 50/60/Hz
Power Consumption max. 500w
Latest Generation Venturi Pump System
Vacuum (Linear, Dual Linear and Fix Control) 0-600 mm Hg.
Aspiration control by Irrigation IV pole and Air Force Infusion. Bottle Height 40-140cm.
Pulsation Frequency
(Adjustable) 1-250 PPS (pulse per second)
Vacuum Response (Venturi Effect, Adjustable) max 0.6-5s
Venturi Pump
Vacuum (Linear, Dual Linear and Fix Control) 0-600 mm Hg.
Vacuum Response (Venturi Effect, Adjustable) max 0.6-5s
Phacoemulsification with Microincision Cataract Surgery (MICS) 1.8mm technology ready
Ultrasound
Frequency 2.5 kHz nominal
Power Output (Linear, Dual Linear and Fix Control) 0-100%
Pulsation (Adjustable)
Continuous ultrasound mode- Frequency 28.5 kHz; Range=0%-100%power in 1% increments
Continuous ultrasound mode- Frequency 2.5 kHz; Range=0%-100%power in 1% increments
Pulse Ultrasound Mode-Frequency 28.5 kHz; Range=1 to 250 pulses per second; Duty Cycle=5 to 95% in 1% increment
Single Burst Ultrasound Mode-Frequency 28.5 kHz; Range=80 to 600msec. Burst with
Fixed Pulse Ultrasound Mode-Frequency 28.5 kHz
Pulse time (Adjustable) 5 to 95% in 1 % increment; 1-250 pulses per second; Duration 2 to 600 msec.
Anterior Vitrectomy
Cutting Rate (Linear, Dual Linear and Fix Control) Pneumatic 0-800 cuts/min
Bipolar Diathermy
Bipolar Wet Field Coagulation
Power Output (Linear and Fix Control)
Liner Mode
Output Range: 7.5 watts Nominal @ 100 ohms
Frequency: 1Mhz nominal Maximum Peak Open Circuit Voltage=120V
Range: Programmable from 0% to 100% in 1% increments
Control: Linear control of coagulation power via the footpedal
Fixed Control
Output Range: 7.5 watts Nominal @ 100 ohms
Frequency: 1Mhz nominal Maximum Peak Open Circuit Voltage=120V
Range: Programmable from 0% to 100% in 1% increments
Control: On/Off control via footpedal
Pulsation (Adjustable) 1 Mhz nominal Maximum Peak Open Circuit Voltage=120V
Computer unit:
19" LED touch screen
Two USB ports for data transfer
Wired or wireless foot control
with remote control
Accessories:
2 pcs. Stellaris Ultrasound Handpiece, Titanium Lightweight with 6 crystals
6 pcs. Microflow needle 30 degrees
Cassettes with tubings 6 packs
Infusion sleeves blue 10 pcs.
Daisy needle wrench
Test Chamber
I/A Handpiece with de-attacheable tips (.04mm)
1 Diathermy Cable
Bipolar forcep
Anterior Vitrectomy Cutter
Standard Requirements:
With Current and valid Certificate of Manufacturer''s compliance with ISO
With Certification to provide Users Manual in English Language
-Operation Manual 2 copies
-Service Manual 2 copies
The brand is more than 10 years in the LOCAL market
Training requirements
Should provide the following Medical Training to (2) end-users (LOCAL)
-Basic Set up and operation training
-Basic Maintenance Training
After Sale Service
Should provide and make available all the service Parts & Accessories for the next five (5) years after the warranty period
Warranty Certificate
One (1) year warranty from the date of delivery when properly installed, maintained and used for its intended purpose and in accordance with Bausch and Lomb Phils., Inc. Instruction. The warranty includes one (1) year service warranty upon acceptance by the end user.
Biometry
Specifications:
Mini PC
3D Mouse Controller
Biometry Probe
Probe Holder
Integrated Test Block
User Manual
Colored Printer
Praeger Shell
Features:
Immersion and Contact Technique
Transportable: Weighing 125 grams
Scleral echo automatically eliminates optic nerve scans
Operates under a PC Windows environment
Can be controlled using a 3D Mouse Controller
Specifications:
Gain: 20 to 110dB
TGC: 0 to -30db
Probe:
Frequency: 11MHz
Diameter of Tip: 5mm (0.20")
Aiming Beam: LED or laser pointer (option)
Depth: 60mm (2.36")
Points on X: 1536
Levels on Y: 256
Axial length measurement:
Types of measurement: anterior chamber, lens, vitreous, total length, automatic calculation of the standard deviation and average, results analysis
No. of measurement: 10
Precision: A+/-0.03mm (A+/-0.002")
Velocity: Adjustable for each segment
Frozen Image: Auto, Auto + save, manual (foot pedal) Automatic detection of scleral spike
Formulas included:
6 Formulas for implant: SRK-T, SRK II, HOLLADAY, BINKHORST-II, HOFFER-Q, HAIGIS
Post-op refractive calculation: Pre-op and Post-op refraction, Pre-op and Post-op Keratometry
: 6 different methods for keratometric correction and implant calculation: history derived, refraction derived, contact lens method, ROSA regression,
SHAMMAS regression, Double K/SRK-T (Dr. ARAMBERRI''s formula)
: 9 different implant powers for desired ametropia value Screen display of 4 different implant calculations implant values available in increments of: 0.25D or 0.50D
Pre-bid Conference
DateTimeVenue
17/11/201510:00 AMBAC Office 2nd floor Provincial Capitol Bldg., Trece Martirez City Cavite Philippines 4109
Documents
Tender Notice