FEDERAL ENVIRONMENT AGENCY has floated a tender for Delivery, Installation, Installation and Briefing on Site: Microtite Plates Reader. The project location is Germany and the tender is closing on 27 Feb 2024. The tender notice number is Z 1.5 - 08 413/0724; Projekt Nr. 188033, while the TOT Ref Number is 96408074. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Germany

Summary : Delivery, Installation, Installation and Briefing on Site: Microtite Plates Reader

Deadline : 27 Feb 2024

Other Information

Notice Type : Tender

TOT Ref.No.: 96408074

Document Ref. No. : Z 1.5 - 08 413/0724; Projekt Nr. 188033

Competition : ICB

Financier : Self Financed

Purchaser Ownership : Public

Tender Value : Refer Document

Purchaser's Detail

Purchaser : FEDERAL ENVIRONMENT AGENCY
Federal Environment Agency, presentation Z 1.5, central award office
Street, house number: Wörlitzer Platz 1
Postcode (post code): 06844
Location: Dessau-Roßlau
Telephone: +49 340 21032350
Fax: +49 340 21042350
Germany
Email :marion.thieme@uba.de
URL :http://www.umweltbundesamt.de/ausschreibungen-zuwendungen

Tender Details

Delivery, installation, installation and briefing on site: microtite plate reader

Extract of technical requirements:

Multi-detection microplatte reader for applications with microtite plates in 6-, 12-, 24-, 48-, 96 and 384 well format from various manufacturers

Detection methods (end point and kinetic for fluorescence and absorption):

Monochromator -based optics (double monochromator system) with free wavelength choice for absorption and fluorescence

Selection of wavelength without a filter change

Accuracy of the wavelength in absorption measurements of plus/minus 0.3 nm

Scan module for individual wells for absorbing absorption, excitation and emission spectra over the entire wavelength range

Top and Bottom-Reading possible

Automated path length correction

Temperage

User -defined shaking functions (linear, orbital), in time and speed variable

Measurement of the smallest volumes

Option to choose several measuring points per Well

Absorption measurement and fluorescence measurement must be possible in a measurement cycle one after the other

z axle optimization for fluorescence measurements

z axis focus for different ...