UltraHighSensitivity and HighSpeed Mass Spectrometry System 1set
Classification: 0024 Professional/Scientific & Controlling Instruments & Apparatus
Summay of Notice: ⑴ Classification of the products to be procured : 24
⑵ Nature and quantity of the products to be purchased : Ultra-High-Sensitivity and High-Speed Mass Spectrometry System 1set
⑶ Type of the procurement : Purchase
⑷ Basic requirements of the procurement :
A Mass Spectrometer
a Must consist of a high-resolution mass analyzer based on electrostatic field Fourier trans form and a high-resolution mass analyzer using an asymmetric ion mirror.
b Must support electrospray ionization (ESI).
c Must have a function to automatically control the ion population inside the mass spectrometer.
d The mass range must include m/z 40
6, 000 for the electrostatic field Fourier transform analyzer and m/z 40
6, 000 for the asymmetric ion mirror analyzer.
e Scan speed must be at least 40 Hz for the electrostatic field Fourier transform analyzer and at least 270 Hz for the asymmetric ion mirror analyzer.
f The electrostatic field Fourier transform analyzer must have a resolution of at least 480, 000 (FWHM) at m/z 200. The asymmetric ion mirror analyzer must have a resolution of at least 80, 000 (FWHM) at m/z 524, and at least 100, 000 (FWHM) at m/z 138 when switching modes.
g Must have an ion source compatible with a nano LC column that integrates emitter, column, and column heater.
B Nano flow Liquid Chromatograph
h The wetted parts must be made of non-metallic, biocompatible materials suitable for proteomics analysis.
i Must support binary high-pressure gradients.
j System pressure tolerance must be at least 150 MPa.
k The nano flow pump must support a flow rate range from 1 nL/min to 100μL/min.
l Must include a degasser capable of online degassing of mobile phases.
m The autosampler must accommodate at least four well plates.
n The sample temperature control range must include 4
40°C....