Procurement Summary
Country: USA
Summary: Stopped-Flow Spectrophotometer System
Deadline: 02 Aug 2017
Posting Date: 19 Jul 2017
Other Information
Notice Type: Tender
TOT Ref.No.: 14987070
Document Ref. No.: RFQ-OH-17-00171
Financier: Self Financed
Purchaser Ownership: -
Tender Value: Refer Document
Purchaser's Detail
Name: Login to see tender_details
Address: Login to see tender_details
Email: Login to see tender_details
Login to see detailsTender Details
Notice type: pre-solicitation
synopsis: stopped-flow spectrophotometer system
naics code: 334516
product service code: 6640
the u.s. Environmental protection agency (epa), office of research and development s (ord), national risk management research laboratory (nrmrl), water systems division (wsd) has a requirement for a stopped-flow spectrophotometer system (sfss) for performing in-house research. The sfss is a fully automated system, which allows absorbance, fluorescence, or conductivity to be used to determine fast reaction kinetics.
The stopped-flow spectrophotometer system must meet or exceed all of the following minimum specifications:
1. Single and double mixing stopped-flow
a fully automated, computer controlled stopped-flow system capable of absorbance and fluorescence measurements, single mixing, and double (sequential) mixing using three reagents and an age loop to incubate the first mixed solution for a set time (minimum aging time of 15 milliseconds or less and a maximum aging time of 1, 000 seconds or greater) before reacting with a third reagent in the observation cell. The stopped-flow must be capable of a dead time of 1.2 milliseconds (ms) or less with a standard cell and 0.5 ms or less with a low volume cell.
-must provide 3-4 syringes for sequential mixing.
-must allow for asymmetric mixing, with ratio mixing up to at least 10:1.
-must provide standard cell with dead time of 1.2 ms or less.
-must provide cuvette geometry able to accommodate multiple detectors.
-must provide pneumatic syringe system for fluid delivery.
-must operate in a working temperature ranging from 0ac to 60ac.
-must provide connectors compatible with water bath systems.
-low volume cell for fast kinetics, up to 2000 s-1
-must provide easily changeable low volume cell for fast kinetics with a dead time of 0.5 ms or less.
3. Multi-wavelength detector
-must provide multi-wavelength detector with variable integ...