Procurement Summary
Country: USA
Summary: Sources Sought Notice of Intent to Sole Source Dual Optical Frequency Comb System
Deadline: 18 Jun 2026
Posting Date: 05 Jun 2026
Other Information
Notice Type: Tender
TOT Ref.No.: 142672583
Document Ref. No.: NIST-SS26-06
Competition: ICB
Financier: Self Financed
Purchaser Ownership: Public
Tender Value: Refer Document
Purchaser's Detail
Name: Login to see tender_details
Address: Login to see tender_details
Email: Login to see tender_details
Login to see detailsTender Details
NIST-SS26-06
6640 / 334516
Primary POC: Liz Timberlake ([email protected])
Secondary POC: Donald Collie ([email protected])
Title: Dual Optical Frequency Comb System
BACKGROUND
***THIS IS A COMBINED SOURCES SOUGHT NOTICE AND NOTICE OF INTENT TO SOLE SOURCE***
The National Institute of Standards and Technology (NIST) is seeking information from sources that may be capable of providing a dual optical frequency comb system. If no alternate sources are identified, the Government intends to issue a Sole Source Award to X3 Optics LLC (1880 S Flatiron Ct, Suite F, Boulder, CO 80301) under the authority of FAR 12.102(a) only one reasonably found source. The North American Industry Classification System (NAICS) code for this acquisition is 334516.
NIST has a need for a dual optical frequency comb system that meets or exceeds the following draft minimum specifications:
The Contractor shall provide one (1) array spectroradiometer that meet the following minimum specifications:
1.Comb Architecture: The system should have two optical frequency combs with 50 MHz repetition rate (or less) each and a controllable difference in repetition rates (from 0 to 10 kHz tuning).
2.Spectral Coverage and Power: The system should have a center wavelength at 1550 nm and 30 nm of bandwidth. Average optical power should be 10 mW (or more).
3.Carrier offset frequency: The system should have an integrated carrier offset frequency detection with high signal to noise ratio of ( 30 dB). The system should have a locked carrier envelope offset frequency with
4. Repetition rate feedback: The system should have both fast repetition rate feedback (100 kHz, or more) and slow feedback ( 10 Hz).
5. Optical or RF lock: The system should have an option to lock a comb line to an optical reference CW laser at 1550 nm. In addition, the system should have an outputting repetition rate beat note for directly locking the repletion rates of eac...
Notice ID: nist-ss26-06
Department/Ind. Agency: commerce, department of
Sub-tier: national institute of standards and technology
Office: dept of commerce nist
Product Service Code: 6640 - laboratory equipment and supplies
NAICS Code: 334516 - Analytical Laboratory Instrument Manufacturing
Inactive Dates: jul 03, 2026
Inactive Policy: 15 days after response date
Documents
Tender Notice