Tender Details
OPQR has a requirement to obtain an advanced dynamic light scattering detector to enhance characterization capabilities for emerging pharmaceutical products and complex biologics. For a quoter's detector to be deemed acceptable, the following salient characteristics shall be met:
QELS Dynamic Light Scattering Detector: Real-time hydrodynamic radius measurement capability that operates inline with Field-Flow Fractionation separation to provide simultaneous size characterization as samples elute from the FFF system.
Compatibility with Existing FFF System: Must be fully compatible with and integrate seamlessly into the existing FFF system, including hardware connections, software integration with chromatography software, and synchronized data acquisition with existing MALS detectors.
High-Sensitivity Detection: Capable of measuring hydrodynamic radii across a wide dynamic range from approximately 10 nm to 1000 nm to accommodate diverse sample types including proteins, protein aggregates, liposomes, and lipid nanoparticles.
Orthogonal Size Validation: Provides diffusion-based size measurements that complement separation-based FFF analysis, enabling validation of particle size distributions and detection of sub-populations or aggregates that may not be fully resolved by separation alone.
Data Correlation and Analysis: Integrated software functionality that enables direct correlation between molecular weight data from MALS detection and hydrodynamic size from DLS detection to provide comprehensive analysis of absolute molar mass, root-mean-square radius, and hydrodynamic radius.
Data Output: Generates validated, audit-trail-compliant data files suitable for regulatory submissions and scientific publications, with full integration into existing software platform for unified data processing and reporting.
Maintenance and Support: Technical support, software updates, preventive maintenance, and alignment with FDA's quality assurance requirements f...