Procurement Summary
Country: Azerbaijan
Summary: Purchase of Machinery and Equipment - Medical Devices and Tools
Deadline: 11 Aug 2025
Other Information
Notice Type: Tender
TOT Ref.No.: 122981139
Document Ref. No.: 115329
Financier: Self Financed
Purchaser Ownership: Public
Tender Value: AZN 1278009.90
Purchaser's Detail
Name: Login to see tender_details
Address: Login to see tender_details
Email: Login to see tender_details
Login to see detailsTender Details
Dermatom, Dermatom device should provide accurate and secure skin cutting function. The device must be available to adjust the depth and width of the cut, parameters to suit different skin types should be offered. Dermatom's cutting knives should be easily changed and high quality should be made of materials. The device's processing speed must be adjusted in a wide range, as well as a stable and smooth cut .Dermatom should be in portable and light design, ensure comfort during use. All components of the device must match sterilization processes and should be made of continuous materials for long-term use. Security mechanisms should be available, including an automatic stop function in the immediate accident., Qty- 1, measurement unit --Dad
Bis, monitor in stationary conditions, resuscitation, ambulance, intensive therapy departments intended for use. It must be suitable for adults, children and newborns.
The device must be equipped with a full touch (Full Touch Screen) color TFT screen.
The screen should be able to be reflected at least 8 waves at the same time.
Should support the following paramets:
5 to take ECG, SPO2, Non-invasive pressure (NIBP), temperature (2 output), respiration,
Ability to connect to the network - RJ45, USB, VGA, RS232, LAN
ECQ
5 Apamali ECG: I, II, III, AVR, AVL, AVF, V
Sensitivity: 1.25mm / mv (x 0.12), 2.5mm / mv (x 0.25), 5mm / mv (x 0.5), 10mm / mv (x 1), 20mm / mv (x 1), 40 mm / mv (x 2), 40 mm / mv (x 4), signal automatically Strengthening function
ECG Opening: 6.25mm / s, 12.5mm / s, 25mm / s, 50mm / s
Digital conversion accuracy from analog: minimum 24 bits
Electrical devices protection: cutting mode 300W, coagulation mode 100w, recovery period ≤ 10s
Monitor Due to 200 different types of arrhythmias in adults and children The detection must have a Detection of St. Segment analysis, heart rate detection and pulse calculation function.
Must have a Pace-Maker detection function on...
Number of bidders required: 3
Opening date and time of envelopes: 11.08.2025 13:00
Date and time of recent submission of proposals: 2025-08-11