Percutaneous Coronary Intervention (PCI) / Two (2) Intra - Aortic Balloon Circulatory Assist Units. The Department of Cardiology at William Beaumont Army Medical Center (WBAMC) requires two Intra-Aortic Balloon Circulatory Assist Units to perform any cardiovascular procedure that requires a readily available mechanical support device. This equipment uses a balloon catheter placed in the aorta to improve heart function/flow and helps the heart pump more blood. WBAMC CATH LAB historically performs approximately 6 high-risk Percutaneous Coronary Intervention (PCI) cases at WBAMC per month that require a mechanical support device like the Intra-Aortic Balloon Pump (IABP). See attached solicitation.
1. ESSENTIAL CHARACTERISTICS:
Seller must provide system that has been approved by the Food and Drug Administration (FDA) and must not have any unresolved FDA recalls.
Seller shall provide the most current system.
System must be portable stand-alone device.
Application must be for adults use only.
System mode of operation must have automatic and manual capabilities.
System display must have high resolution, color touch screen monitor with screen size up to 14” with parameters readings for three trace, Electrocardiogram (ECG), arterial pressure, balloon pressure, trigger mode, operational mode, assisted and unassisted pressure values.
System must have balloon volume adjustment with increments of 0.5cc.
System must have timing adjustment increments of 1 second.
System must have signal outputs of ECG, Arterial pressure, fluid filled transducer, fiberoptic sensor.
System must have the capabilities of auto condensation removal without interruption.
System must alarm notifications for trigger loss, vacuum pressure loss, balloon disconnect, balloon leak, balloon overinflate, systolic inflate, loading failure, high pressure, power loss, device malfunctions, low battery indicator and temporary alarm silencing by end-user.
System must...
Notice ID: w81k0026qa153
Department/Ind. Agency: dept of defense
Original Set Aside: service-disabled veteran-owned small business (sdvosb) set-aside (far 19.14)
Sub-tier: dept of the army
Office: w40m mrc0 west
Product Service Code: 6515 - medical and surgical instruments, equipment, and supplies
NAICS Code: 339112 - Surgical and Medical Instrument Manufacturing
Inactive Dates: jun 04, 2026
Inactive Policy: 15 days after date offers due
Documents
Tender Notice
Documents.zip