Procurement Summary
Country: USA
Summary: Fiber Optic Hydrophone
Deadline: 27 Jul 2016
Posting Date: 13 Jul 2016
Other Information
Notice Type: Tender
TOT Ref.No.: 7289927
Document Ref. No.: RFQFDA1168701
Financier: Self Financed
Purchaser Ownership: -
Tender Value: Refer Document
Purchaser's Detail
Name: Login to see tender_details
Address: Login to see tender_details
Email: Login to see tender_details
Login to see detailsTender Details
Rfqfda1168701
part 1 - introduction july 12, 2016
the food and drug, (fda) has a requirement to purchase a fiber optic hydrophone, onda product # hfo-690, brand name or equal.
This is a combined synopsis/solicitation for commercial items prepared in accordance with the format in subpart 12.6, (streamlined procedures for evaluation and solicitation for commercial items) as supplemented with additional information included in this notice. This announcement constitutes the only solicitation.
The solicitation is issued as request for quote (rfq), using the simplified acquisitions procedures of far 13. The solicitations document/incorporated provisions and clauses are those in effect through the current federal acquisition circular: fac 2005-88.
The associated north american industry classification system (naics) code is 334511: search, detection, navigation, guidance, aeronautical, and nautical system and instrument manufacturing, small business size standard is 750 employees. This acquisition is set aside for small business.
Contract type: commercial item: firm fixed price, fob destination
part 2 - prices/cost
clin description
unit price qty unit of issue price
0001 fiber optic hydrophone, omda product #hfo-690 brand name or equal, laser source 690nm, set of 5 fiber optic probes (2 meters in length each), fiber cleaver, fiber stripper tool, optical probe holder, self-calibrating, deconvolution software, protective eyewear,
delivery included.
$ 1 ea $
fob destination
estimated delivery date: 8/15/2016
part 3 -statement of work
fiber optic hydrophone system requirement:
can support acoustic pressure up to 500 mpa
frequency response from 3 khz to 150 mhz
spatial resolution of at least 100 microns
sensitivity of at least 6 mv/mpa
noise equivalent pressure no more than 0.25 mpa (2mhz bw), 0.3 mpa (10 mhz bw), and 0.5 mpa (100 mhz bw)
temperature sensitivity no more than 10 mv/degree
one year warranty on all parts and labor.
Part 4 - inspection and acceptance:
-c the contractor s officer s representative (cor) will perform inspection and acceptance of the scientific instrument.