DEPARTMENT OF VETERANS AFFAIRS has floated a tender for Excimer Laser System And Disposable Catheters For The Treatment Of Peripheral And Coronary Arterial Disease. The project location is USA and the tender is closing on 06 Mar 2019. The tender notice number is 36C26119Q0374, while the TOT Ref Number is 31137911. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : USA

Summary : Excimer Laser System And Disposable Catheters For The Treatment Of Peripheral And Coronary Arterial Disease

Deadline : 06 Mar 2019

Other Information

Notice Type : Tender

TOT Ref.No.: 31137911

Document Ref. No. : 36C26119Q0374

Competition : ICB

Financier : Self Financed

Purchaser Ownership : -

Tender Value : Refer Document

Purchaser's Detail

Purchaser : DEPARTMENT OF VETERANS AFFAIRS
Contracting Office Address: Department Of Veterans Affairs
Va Sierra Pacific Network (Visn 21)
Va Pacific Islands Health Care System
459 Patterson Rd
Honolulu Hi 96819-1522

Primary Point of Contact: Moneque Rodriguez
USA

Tender Details

This is a sources sought for an excimer laser with fda clearance for use in multiple minimally invasive cardiovascular procedures. This excimer laser shall have the ability to use cool ultraviolet energy to ablate arterial blockages and remove pacemaker leads. This notice allows for vendors to name brand name or equal products. All items shall be new, and not refurbished or recycled. Capabilities statements shall include the name of company, business title, e-mail address and phone number, company s business size, gsa/sewp/fss contract number (to include expiration date), duns number, and small business status as verified from vetbiz.gov. All vendors must be properly registered in the system award management database (sam) and vetbiz.gov. Capability statements are due no later than march 06, 2019 3 pm est.





Clin 0001



description: excimer laser system: cvx-300



qty: 1



unit price: $



extended price: $



total price: $



brand name or equal





technical specifications/salient characteristics



compact and portable



must have 220 ac power supply



reliable and durable



lcd full-color display



real time use for increased accuracy, treatment, and time savings



include adjustable height roll stand with quiet wheels



shorter case times for decreased radiation and contrast exposure



start up time is 5 minutes from the time it is turned on to when the physician can start using it.



Device can be wheeled between rooms to be utilized in more than one cath lab.



Power cable no less than 10 feet in length



length no more than 49 inches. Height no more than 35 inches. Width no more than 25 inches.



Xecl pulsed excimer laser system



wavelength: 308 nm



pulse width: 125-200 ns fwhm



catheter output fluence: 30-80 ml/mm2



repetition rate rang: 25 80 hz



weight: approx. 650 lbs



set-up accessories



end-user training

please consult the list of document viewers if you cannot open a file.

Attachment

type: other (draft...

Documents

 Tender Notice