This is a sources sought notice only. This is not a solicitation for bids, proposals, proposal abstracts, or quotations. The purpose of this sources sought notice is to obtain information regarding the availability and capability of all qualified sources to perform a potential requirement. The responses received from interested contractors will assist the government in determining the appropriate acquisition method.
The department of veterans affairs (va), network contracting office (nco) 20, is conducting market research to identify potential vendor information pages (vip) verified (www.vip.vetbiz.gov) service disabled veteran owned business (sdvosb) or veteran owned business (vosb) sources which can provide the following equipment:
chlorine analyzers features
o epa 334.0 compliant
o reagent free measuring
o complete panel system allows for quick and easy installation
o built-in flow regulator maintains constant flow and pressure to the sensors regardless of inlet pressure
o pre-wired panel includes a 100/240 vac power supply, two 4 to 20 ma outputs and two mechanical relays
o optional automatic ph compensation
" general
o compatible 3-2630-1 free chlorine electrode, 0 to 2 ppm / 3-2650-7 amperometric electronics
o 3-2630-2 free chlorine electrode, 0 to 5 ppm / 3-2650-7 amperometric electronics
o 3-2630-3 free chlorine electrode, 0 to 20 ppm / 3-2650-7 amperometric electronics
o signet 3-2724-00 flat ph electrode, 0 to 14 ph/ 3-2750-7 ph sensor electronics
" materials
o panel black acrylic
o flow cell acrylic
o wiring enclosure polycarbonate
" wetted materials
o flow cell, spacer rings acrylic
o flow regulator housing polycarbonate
o strainer, e-clip, regulator spring, float stainless steel
o valves, vent polypropylene
o flow cell o-rings, diaphragm epdm, fkm
o chlorine electrode pvc, pvdf, fpm
o ph electrode pps, glass, uhmw pe, fpm
o sealing tape on valves, plug and vent ptfe
o plug polyethylene
" max. Temperature/pressure rating
o system inlet pressure rating 1 to 8 bar 15 to 120 psi
o pressure regulator 0.69 bar (10 psi) variation over all ranges of flow and pressure
o flow tolerance a+/- 15% or rated specification above
o flow rate limits 30.24 to 45.36 lph 8 to 12 us gal/h
o storage temperature 0 ac to 65 ac 32 af to 149 af
o operating temperature 0 ac to 45 ac 32 af to 113 af
" electrical
o ac input - standard configuration 100 to 240 vac nominal 50 to 60 hz, 0.17 a at 100 vac
o dc input - optional configuration 12 to 24 vdc a+/-10% regulated, 250 ma max.
" environmental
o relative humidity 0 to 95%
o maximum altitude 2000 m (6, 562 ft)
o enclosure nema 4x (with output wire glands sealed)
potential sd/vosb s having the capabilities necessary to provide the above stated equipment are invited to respond to this sources sought notice via e-mail to craig wagstaff at [email protected] no later than september 19, 2016. no telephone inquiries will be accepted.
responses should include the following information: company name, address, and business size; point of contact name, phone number, and e-mail address; suggested equipment information including manufacturer, model and delivery schedule; whether system is presently offered on a current gsa federal supply schedule contract. naics code 334516 is applicable to determine business size standard. any questions or concerns may also be directed to craig wagstaff via e-mail.
disclaimer and important notes: this sources sought notice does not obligate the government to award a contract or otherwise pay for the information provided in response. the government reserves the right to use information provided by respondents for any purpose deemed necessary and legally appropriate. the government will treat any information received as proprietary and will not share such information with other companies. any organization responding to this sources sought notice should ensure that its response is complete and sufficiently detailed to allow the government to determine the organization s qualifications to perform the work. respondents are advised that the government is under no obligation to acknowledge receipt of the information received or provide feedback to respondents with respect to any information submitted. the government may or may not issue a solicitation as a result of this announcement. there is no solicitation available at this time. please consult the list of document viewers if you cannot open a file.
attachment
type: other (draft rfps/rfis, responses to questions, etc..) label:
attachment
posted date: september 14, 2016 https://www.vendorportal.ecms.va.gov/fbodocumentserv...
https://www.vendorportal.ecms.va.gov/fbodocumentserver/documentserver.aspx?documentid=3002868&filename=va260-16-n-0958-000.docx
to copy full link address, right-click and select "copy shortcut" or "copy link location."description: va260-16-n-0958 va260-16-n-0958.docx
boise va medical center;500 w. fort street;boise, id
83702
usa
52 -- measuring tools
naics code:
334 -- computer and electronic product manufacturing/334516 -- analytical laboratory instrument manufacturing
334 -- computer and electronic product manufacturing/334516 -- analytical laboratory instrument manufacturing
for help: federal service deskaccessibility
added: sep 14, 2016 11:04 am
Documents
Tender Notice