SOUTHERN TAGALOG REGIONAL HOSPITAL has floated a tender for Calibration and Preventive Maintenance of Mobile X-Ray Machine. The project location is Philippines and the tender is closing on 14 May 2024. The tender notice number is SV-2024-026B, while the TOT Ref Number is 100899625. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Philippines

Summary : Calibration and Preventive Maintenance of Mobile X-Ray Machine

Deadline : 14 May 2024

Other Information

Notice Type : Tender

TOT Ref.No.: 100899625

Document Ref. No. : SV-2024-026B

Competition : ICB

Financier : Self Financed

Purchaser Ownership : Public

Tender Value : PHP 150000.00

Purchaser's Detail

Purchaser : SOUTHERN TAGALOG REGIONAL HOSPITAL
Raiza Yanga Ralleca
BAC - Secretariat
Brgy. Habay II, Bacoor City
Bacoor
Cavite
Philippines 4102
63-966-0888842
Philippines
Email :strhbacsec@gmail.com

Tender Details

Tender are invited for Calibration and Preventive Maintenance of Mobile X-Ray Machine

Procurement Mode : Negotiated Procurement - Small Value Procurement (Sec. 53.9)

Classification : Goods

Category : Hospital / Medical Equipment Services

Contract Duration : 0 Day/s

Budget for the Contract : PHP 150, 000.00

Area of Delivery : Cavite

Description : Calibration and Preventive Maintenance of Mobile X-Ray Machine



Mobile X-Ray Machine

a. Frequency: SEMI-ANNUAL

b. Quantitative Test (Calibration Parameter):

(Calibration must include but not limited to the ff test)

b.1 Mechanical Checks

• Collimator Assessment

b.2 Generator Checks

• Overload Circuit Protection

• kVp Accuracy

• mA Linearity

• Exposure Time

b.3 Flat Panel Detector

b.4 System Checks



Brand/Model: SHIMADZU/MOBILEART EVOLUTION

1. MPE09FD56001

Brand/Model: ANGELL/DP326B-3

1. DX0362020129701



Standard Requirements:

a. Supplier must provide Service and Test Report for each unit as compliance for the requirement of STRH. Details of Calibration Tools such as brand, model, serial number, and calibration date range must be indicated in either in Service Report or Test Report.

b. Supplier must provide a copy of Calibration Report or Certificate for the Calibration Tools used.

c. Supplier must provide a calibration sticker per unit indicating the start and end date of calibration.

d. Supplier must conduct ...