This requirement is being acquired as a brand-name-or-equal procurement. Equal products meeting all stated salient characteristics will be considered.
The instrument shall be capable of delivering a minimum volume of 250ml and a maximum volume of at least 1000ml.
The instrument shall have a dispensing accuracy of ±5mL or less.
The instrument shall be equipped with in-line heating that is capable of heating dissolution media to at least 45°C.
The instrument shall have a temperature accuracy of ±2°C or better.
The instrument shall be capable of degassing media up to a maximum of 3.0ppm.
The instrument shall be compatible with media containing up to a maximum of 2% surfactants.
The instrument shall have an onboard controller and user interface to select stored programs and set parameters for manual use.
The instrument shall be capable of storing programmed methods and running in automatic or manual mode when preparing dissolution media.
The instrument shall be capable of storing dispensing reports that include the volume and temperature dispensed.
The instrument shall be portable with a cart or other mechanism, included, to allow the instrument to be moved to each dissolution instrument for use.
The instrument shall be equipped with a dispensing nozzle that allows for dissolution vessels to be filled in place, while on the dissolution apparatus.
Electrical requirement must be compatible with a standard US outlet, 115V, 50 or 60 Hz.
Safety Requirements
The Contractor shall ensure all equipment complies with applicable laboratory safety standards and manufacturer safety requirements. All installation activities shall be performed in accordance with applicable safety regulations to prevent hazards to personnel and facilities.
Trade and Service Specifications
The instrument must be a newly manufactured unit, not used and refurbished or previously used for demonstration.
FOB Point destination to inc...
Notice ID: ssn_oc_2026_133856
Department/Ind. Agency: health and human services, department of
Original Set Aside: total small business set-aside (far 19.5)
Sub-tier: food and drug administration
Office: fda office of acq grant svcs
Product Service Code: 6632 - analysis instruments, medical laboratory use
NAICS Code: 334516 - Analytical Laboratory Instrument Manufacturing
Inactive Dates: may 20, 2026
Inactive Policy: manual
Documents
Tender Notice
Documents.zip