LEIBNIZ-INSTITUT FÜR GEMÜSE-UND ZIERPFLANZENBAU GROßBEEREN/ERFURT E. V. has floated a tender for Contract notice: Mass spectrometer. The project location is Germany and the tender is closing on 02 Nov 2018. The tender notice number is 407678-2018, while the TOT Ref Number is 26909556. Bidders can have further information about the Tender and can request the complete Tender document by Registering on the site.

Expired Tender

Procurement Summary

Country : Germany

Summary : Contract notice: Mass spectrometer

Deadline : 02 Nov 2018

Other Information

Notice Type : Tender

TOT Ref.No.: 26909556

Document Ref. No. : 407678-2018

Competition : ICB

Financier : Self Financed

Purchaser Ownership : -

Tender Value : Refer Document

Purchaser's Detail

Purchaser : LEIBNIZ-INSTITUT FÜR GEMÜSE-UND ZIERPFLANZENBAU GROßBEEREN/ERFURT E. V.
Office Name: Leibniz-Institut für Gemüse-und Zierpflanzenbau Großbeeren/Erfurt e. V.
Address: Theodor-Echtermeyer-Weg 1
Town: Großbeeren
Postal Code: 14979


Phone: +49 33701-78114
Fax: +49 33701-55391
Germany
Email :vw-ltg@igzev.de
URL :www.igzev.de

Tender Details

Object of the contract
Mass spectrometer

Description: Contract notice: Mass spectrometer

Authority Type: Other
Contact Nature: Supplies
Procedure: Open procedure
Document: Contract notice
Regulation: European Union
Award criteria: The most economic tender
CPV code: 38433100, 38433100
CPV Description: Mass spectrometer.

Mass spectrometer

Reference Number: 2018_LC-MS / MS System

LC-MS / MS System

Main Site: Leibniz Institute of Vegetable and Ornamental Plants Grossbeeren / Erfurt e. V. Theodor-Echtermeyer-Weg 1 14979 Grossbeeren

LC-MS / MS System

With the following requirements for the source

- API source with heating of the LC columns up to 400C and support of exchangeable ion sources: ESI, APCI, DART, Nanospray,

- ESI compatible with flow rates from 1 uL / min to 1 mL / min, without splitting,

- APCI compatible with flow rates from 50 uL / min to 2 mL / min, without splitting,

- Nanospray compatible with Flow rates from 1 nL / min to 50 μL / min,

- Dual CCD cameras (X, Y position) for fine positioning,

API source with automatic tuning program.

Mass analyzer

- Desktop,

- Dual quadrupole orbitrap-in series,

- mass range: m / z 50-6 000, with continuous ionic isolation with an ion window of 0.4 to 2 450 Da across the entire mass range,

- mass accuracy: 1 ppm RMS for internal and 3 for external calibration,

- Data acquisition modes: full scan, selected ion monitoring scan, product ion scan, parallel reaction monitoring, multiplex SIM and multiplex MS / MS modes with up to-10 Selected Vorlaeuferionen, data independent acquisition, neutral loss scans, retention time scheduled SIM and MS / MS scan,

- mass resolution: 140 000 at m / z 200 for all scan modes and both polarities,

- Linear dynamic range of 5 000: 1,

- polarity change: a spectrum in positive and negative mode in 1s with a resolution of 35 000,

- scan rate: full scan or SIM modewith up to 12 scans per second with a resolution of 17, 500 at m / z 200, MS / MS scan mode at up to 10 scans per second with a resolution of 17, 500 at m / z 200,

- Collision energy profiling for using different collision energies for HCD fragmentation,

- No external cooling, no gases other than nitrogen (nitrogen supply is guaranteed),

- Maximum dimensioning: 95 x 85 x 95 cm.

Nano-UHPLC

- Full compatibility with mass spectrometer,

- Works with nano, capillary and micro flow rates,

- Pressure resistance up to 1000 bar,

- 2-way / 6-port injection valve,

- Sample tempering from 4-45C (up to 22C below room temperature), - - Injection volume from 0-20 μl,

- Autosampler with up to 3 well plates (96 or 384, standard autosampler vials),

- Full and partial loopfeninjection and microliter pick-up,

starter kit, including C18 columns for peptide separation,

- Maximum dimension: 45 x 100 x 55 cm.

Control and evaluation software

- Complete control of the mass spectrometer and the nanoUHPLC under one software,

- Evaluation software supports various search algorithms (SEQUEST, Z-Core, Mascot) and combines the output of different algorithms-for cross-validation of results,

evaluation software supports various fragmentation techniques (CID, HCD, ETD),

absolute quantification by peptide labeling (TMT, iTRAQ), and relative quantification (label-free),

generation of a false discovery rate in protein identification,

- Support of data interpretation by automatic GO annotation, PTM site detection andDatabase Search, Delivery, Installation and Briefing

- No Charge for Delivery, Installation and Briefing

Training

- 2 Days Mass Spectrometer Training, On-Site,

- 2 Days Fluid Chromatography Training, On-Site,

- 1 Day Training for evaluation software, either on site or at the manufacturer,

maintenance

- maintenance contract for 24 months (start of contract after expiry of de12-month warranty, including warranty extension, annual maintenance and software updates),

- labor and travel time costs and spare parts included,

- service on site within 3 working days.

According to minimum requirements of the Brandenburg Procurement Law

Announcement ID: CXP9YRZD45Z


Internet address (URL): www.igzev.de

Documents

 Tender Notice