Procurement Summary
State : Jammu and Kashmir
Summary : Supply and Installation of Gradient Pcr
Deadline : 26 Sep 2019
Other Information
Notice Type : Tender
TOT Ref.No.: 36283804
Document Ref. No. : BGSBU/Bio/19/149/1
Competition : NCB
Financier : Self Financed
Purchaser Ownership : -
Document Fees : Refer Document
Tender Value : Refer Document
EMD : Refer Document
Purchaser's Detail
Purchaser : BABA GHULAM SHAH BADSHAH UNIVERSITY
Address: College of Engineering and Technology Rd, Rajouri, Jammu and Kashmir 185234
Country: India
Email: masgerghazi@gmail.com
Website: www.coetbgsbu.org
Tender Details
Sealed Quotations Are Invited for Supply and Installation of Gradient Pcr. ï‚· the System Should Have 96 Wells with 6 Separate Peltier Blocks with Sample Capacity
96x0.2ml Tubes or 96well Plates.
ï‚· Each Block to Accommodate 16 Wells and Having the Ability to Set up Pcr with A Specific
Temperature Differential of Up to 5 Degree Centigrade Between Blocks.
ï‚· Run up To 6 Separate Temperatures in The Same Plate with User Defined Time to Determine
the Optimal Annealing Temperatures.
ï‚· on Board Tm Calculator Facility to Approximate the Optimal Annealing Temperature.
 Temperature Accuracy ±0.25o C (35–99.9o C)
ï‚· Block Heating/ Cooling Rate: 3.9 O C/sec or More.
ï‚· Temperature Range 0o C to 100o C
ï‚· Temperature Uniformity 0.5 deg c
ï‚· gradient range 0.1o c to 25 o c across the block.
ï‚· vga colour touch screen allowing for easy intuitive graphical user interface
programming.
ï‚· choice of saving the methods up to 800 to the instrument or unlimited.
ï‚· programmable heat lid cover from 50 degree to 105 degree centigrade for efficient pcr
optimization
ï‚· scalability: capability to interlink up to 11 pcr systems via single ethernet hub. auto-
start on power failure
ï‚· portability: the system has a usb port to transfer methods from one machine to another.
ï‚· system should be certified with ce, industrial scientific & medical and emc
ï‚· should have atleast 200 installations of pcr in india.
ï‚· warranty 1 year from the date of installation
ï‚· should be supplied with ups backup.
 should be supplied with uv transilluminator; 20×20 cm; 254/312/365nm wavelength
uv tubes.
Documents